Code | CSB-E14068m |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase in six months till depleted. Apply now |
Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<15% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of mouse CRF in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 95 | ||||
Range % | 90-100 | |||||
1:2 | Average % | 97 | ||||
Range % | 90-105 | |||||
1:4 | Average % | 96 | ||||
Range % | 95-102 | |||||
1:8 | Average % | 94 | ||||
Range % | 90-101 |
The recovery of mouse CRF spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 89 | 85-96 | ||||
EDTA plasma (n=4) | 94 | 87-99 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||||||||||||||||||||||||||||||||||||||
|
This Mouse CRH ELISA Kit was designed for the quantitative measurement of Mouse CRH protein in serum, plasma. It is a Competitive ELISA kit, its detection range is 2.5 pg/mL-40 pg/mL and the sensitivity is 0.675 pg/mL .
Sample type: Serum
Sample species: Mouse
Sample dilution: 1:5
Review: CSB-E14068m kit was used to detect the serum level of CRF in mice. After 5 times of dilution, the normal and model groups and the administration groups were all within the range of standard curve. All experiments were completed within three and a half hours, and the test results were stable and the standard curve was good
By Anonymous