| Code | CSB-E09920h |
| Size | 96T,5×96T,10×96T |
| Price | Request a Quote |
| Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
| Have Questions? | Leave a Message or Start an on-line Chat |
| Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<10% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of human PPIA/CYPA in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:1 | Average % | 92 | ||||
| Range % | 87-95 | |||||
| 1:2 | Average % | 100 | ||||
| Range % | 95-104 | |||||
| 1:4 | Average % | 90 | ||||
| Range % | 85-94 | |||||
| 1:8 | Average % | 98 | ||||
| Range % | 92-101 | |||||
| The recovery of human PPIA/CYPA spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 90 | 84-94 | ||||
| EDTA plasma (n=4) | 95 | 89-98 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This Human PPIA ELISA Kit was designed for the quantitative measurement of Human PPIA protein in serum, plasma, tissue homogenates, cell lysates. It is a Sandwich ELISA kit, its detection range is 3.12 ng/mL-200 ng/mL and the sensitivity is 0.78 ng/mL.
Sample type: Serum
Sample species: Human
Sample dilution: No dilution
Review: I used the Human cyclophilin A (PPIA/CYPA) ELISA Kit. The samples were human serum infected with the virus and healthy people\'s serum, of course, the healthy people have no other chronic diseases. From the experimental results, the kit performance is good, the negative hole value is between 0.50 and 0.07. This kit really has no slots, see the picture
By Anonymous
Sample type: Serum
Sample species: Human
Sample dilution: 1:4
Review: I used the Human cyclophilin A (PPIA/CYPA) ELISA Kit item number (CSB-E09920h) to detect the PPIA protein in human serum and obtained a good standard curve. The human serum could be well detected after being diluted four times.
By Anonymous
Can you provide me with the information if antibodies are polyclonal or monoclonal?