Code | CSB-E11298r |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * The sample kit cost can be deducted from your subsequent orders of 96T full size kits of the same analyte at 1/5 per kit, until depleted in 6 months. Apply now |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of rat BMG in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 87 | ||||
Range % | 82-92 | |||||
1:2 | Average % | 96 | ||||
Range % | 92-101 | |||||
1:4 | Average % | 100 | ||||
Range % | 95-105 | |||||
1:8 | Average % | 90 | ||||
Range % | 85-95 |
The recovery of rat BMG spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 97 | 90-104 | ||||
EDTA plasma (n=4) | 93 | 89-97 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
The Rat β2-microglobulin (BMG/β2-MG) ELISA Kit is your go-to solution for accurate and reliable quantification of Beta-2-microglobulin in rat serum, plasma, urine, and tissue homogenates.
Designed for researchers in the field of immunology、autoimmune diseases and renal function research, this ELISA kit provides a quantitative measurement of B2M levels with a detection range of 0.156 μg/mL-10 μg/mL and a sensitivity of 0.039 μg/mL, ensuring precise and reproducible results.
The assay principle is based on a sandwich measurement, where the captured rat β2-microglobulin is detected using a specific antibody conjugated to horseradish peroxidase. The kit's assay time is 1-5 hours, and the sample volume required is between 50-100ul.
The Rat β2-microglobulin (BMG/β2-MG) ELISA Kit is compatible with standard 96-well microplates and requires a detection wavelength of 450 nm. This ELISA kit has been cited in more than 3 scientific publications, demonstrating its robust performance and reliability.