Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human ANX-Ⅲin various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 95 | ||||
Range % | 92-101 | |||||
1:2 | Average % | 90 | ||||
Range % | 82-96 | |||||
1:4 | Average % | 101 | ||||
Range % | 99-104 | |||||
1:8 | Average % | 87 | ||||
Range % | 82-99 |
The recovery of human ANX-Ⅲ spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 90 | 82-96 | ||||
EDTA plasma (n=4) | 95 | 89-100 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]()
|
This Human ANXA3 ELISA Kit was designed for the quantitative measurement of Human ANXA3 protein in serum, plasma, cell culture supernates, tissue homogenates, cell lysates. It is a Sandwich ELISA kit, its detection range is 0.78 ng/mL-50 ng/mL and the sensitivity is 0.195 ng/mL .
Sample type: Plasma (anticoagulant)
Sample species: Human
Sample dilution: No dilution
Review: I used csb-e12157h for testing clinical blood specimens from sepsis patients, the specimens were not diluted and processed, and the standard curve was good, resulting in an increase in anxa3 in blood specimens from sepsis patients. The kit has good sensitivity and works well.