| Code | CSB-EL005291HU |
| Size | 96T,5×96T,10×96T |
| Price | Request a Quote |
| Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
| Have Questions? | Leave a Message or Start an on-line Chat |
| Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<10% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of human CFP in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:100 | Average % | 95 | ||||
| Range % | 90-100 | |||||
| 1:200 | Average % | 106 | ||||
| Range % | 101-110 | |||||
| 1:400 | Average % | 98 | ||||
| Range % | 92-103 | |||||
| 1:800 | Average % | 90 | ||||
| Range % | 85-97 | |||||
| The recovery of human CFP spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 84 | 80-90 | ||||
| EDTA plasma (n=4) | 93 | 89-98 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This Human CFP ELISA Kit was designed for the quantitative measurement of Human CFP protein in serum, plasma, cell culture supernates, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 12.5 ng/mL-800 ng/mL and the sensitivity is 3.12 ng/mL.
Sample type: Plasma (anticoagulant) EDTA
Sample species: Human
Sample dilution: : 1:200
Review: I used CSB-EL005291HU to detect the CFP level in human plasma. The highest standard point of 3.38 and the lowest standard point of 0.08 were excluded, with r²=0.99989 and the range of 2-100 ug/ml.
By Anonymous