Code | CSB-E08436m |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * The sample kit cost can be deducted from your subsequent orders of 96T full size kits of the same analyte at 1/5 per kit, until depleted in 6 months. Apply now |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of mouse PAP in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:5 | Average % | 87 | ||||
Range % | 83-94 | |||||
1:10 | Average % | 97 | ||||
Range % | 92-102 | |||||
1:20 | Average % | 105 | ||||
Range % | 98-111 | |||||
1:40 | Average % | 93 | ||||
Range % | 87-99 |
The recovery of mouse PAP spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 92 | 88-96 | ||||
EDTA plasma (n=4) | 101 | 95-107 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
This Mouse PAP ELISA Kit was designed for the quantitative measurement of Mouse PAP protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 3.12 pg/mL-200 pg/mL and the sensitivity is 0.78 pg/mL.
There are currently no reviews for this product.
Can you comment on the nature of the antibodies used for the ELISA? Does one of the antibodies exclude the detection of plasminogen?
For CSB-E08436m kit, the target is a compound. There are two different antibodies used in the kit, the capture antibody, and the detection antibody. They are respectively specific to the two parts of the compound and they cannot detect plasminogen.