Call us
301-363-4651 (Available 9 a.m. to 5 p.m. CST from Monday to Friday)
Code | CSB-E08745Rb |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * The sample kit cost can be deducted from your subsequent orders of 96T full size kits of the same analyte at 1/5 per kit, until depleted in 6 months. Apply now |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of rabbit IL-2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 87 | ||||
Range % | 82-93 | |||||
1:2 | Average % | 95 | ||||
Range % | 91-100 | |||||
1:4 | Average % | 86 | ||||
Range % | 80-95 | |||||
1:8 | Average % | 93 | ||||
Range % | 82-98 |
The recovery of rabbit IL-2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 92 | 80-98 | ||||
EDTA plasma (n=4) | 93 | 84-100 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
This Rabbit IL2 ELISA Kit was designed for the quantitative measurement of Rabbit IL2 protein in serum, plasma. It is a Sandwich ELISA kit, its detection range is 31.25 pg/mL-2000 pg/mL and the sensitivity is 7.8 pg/mL.