Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of chicken IL-2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 89 | ||||
Range % | 85-96 | |||||
1:2 | Average % | 90 | ||||
Range % | 85-97 | |||||
1:4 | Average % | 99 | ||||
Range % | 94-106 | |||||
1:8 | Average % | 97 | ||||
Range % | 92-104 |
The recovery of chicken IL-2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 100 | 92-109 | ||||
EDTA plasma (n=4) | 96 | 90-104 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]()
|
The Chicken Interleukin 2 (IL-2) ELISA Kit is used for accurate and precise measurement of IL-2 in samples from Gallus gallus (Chicken) species. It is designed for use with serum, plasma, and tissue homogenates, and provides a detection range of 0.18 pg/mL-12 pg/mL, with a sensitivity of 0.04 pg/mL.
This ELISA Kit is based on the Sandwich assay principle and has a measurement wavelength of 450 nm. With an assay time of just 1-5 hours and a sample volume of 50-100 ul, this kit is both efficient and convenient for laboratory use.
This product is ideal for researchers in the field of Immunology, and has been cited in over 25 scientific publications. With its high sensitivity and accuracy, this kit will provide you with reliable results for your research needs.