| Code | CSB-E14290h-IS |
| Size | 96T,5×96T,10×96T |
| Price | Request a Quote |
| Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
| Have Questions? | Leave a Message or Start an on-line Chat |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human IL-29 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | ||||
1:1 | Average % | 92 | ||||
Range % | 88-97 | |||||
1:2 | Average % | 98 | ||||
Range % | 91-105 | |||||
1:4 | Average % | 89 | ||||
Range % | 84-94 | |||||
1:8 | Average % | 93 | ||||
Range % | 86-99 | |||||
The recovery of human IL-29 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 94 | 89-98 | ||||
EDTA plasma (n=4) | 87 | 82-93 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||
| |||||||
pg/ml | OD1 | OD2 | Average | Corrected | |||
200 | 1.876 | 1.934 | 1.905 | 1.714 | |||
100 | 1.220 | 1.267 | 1.243 | 1.052 | |||
50 | 0.717 | 0.728 | 0.722 | 0.531 | |||
25 | 0.429 | 0.419 | 0.424 | 0.233 | |||
12.5 | 0.341 | 0.321 | 0.331 | 0.140 | |||
6.25 | 0.284 | 0.273 | 0.279 | 0.087 | |||
3.12 | 0.235 | 0.244 | 0.239 | 0.048 | |||
0 | 0.187 | 0.195 | 0.191 |
| |||
This ELISA kit provides quantitative measurement of interleukin-29 (IL-29/IFN-λ1) in human samples, supporting research into type III interferon-mediated antiviral and mucosal immune responses. Because IL-29 is often expressed at low levels in circulation and is transiently induced during viral infection or at epithelial barrier surfaces, detecting it requires an assay calibrated for the picogram-per-milliliter range with validated performance across diverse sample matrices.
Detection Range: The 3.12–200 pg/mL range aligns with reported IL-29 concentrations in human serum and plasma during acute viral infections such as hepatitis C and influenza, where levels typically fall in the low-to-mid picogram range. Most stimulated samples can be measured directly without extensive dilution.
Sensitivity: At 1.56 pg/mL, the kit can detect IL-29 in baseline or early-infection samples where type III interferon induction is minimal, including unstimulated PBMC supernatants or serum from chronically infected patients with dampened IFN-λ responses.
Sample Compatibility: Validation across serum, plasma, and tissue homogenates enables researchers to correlate systemic IL-29 levels with local expression at mucosal sites such as lung, liver, or intestinal tissue — critical for studying compartmentalized antiviral immunity.
Assay Efficiency: The 1–5 hour protocol using only 50–100 µL of sample accommodates limited-volume clinical specimens and time-course studies requiring multiple sampling points from the same donor.
This kit is well-suited for studies of hepatitis C pathogenesis, respiratory viral immunity, and IFN-λ polymorphism-associated treatment response.