Call us
						301-363-4651 (Available 9 a.m. to 5 p.m. CST from Monday to Friday)
					| Code | CSB-E07925m | 
| Size | 96T,5×96T,10×96T | 
| Price | Request a Quote | 
| Trial Size | 24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below. | 
| Have Questions? | Leave a Message or Start an on-line Chat | 
| Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<10% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of mouse Col Ⅲ in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:20 | Average % | 89 | ||||
| Range % | 82-96 | |||||
| 1:40 | Average % | 98 | ||||
| Range % | 94-102 | |||||
| 1:80 | Average % | 97 | ||||
| Range % | 90-104 | |||||
| 1:160 | Average % | 86 | ||||
| Range % | 82-90 | |||||
| The recovery of mouse Col Ⅲ spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 93 | 87-99 | ||||
| EDTA plasma (n=4) | 100 | 96-104 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|   
 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This Mouse Col Ⅲ ELISA Kit was designed for the quantitative measurement of Mouse Col Ⅲ protein in serum, plasma. It is a Sandwich ELISA kit, its detection range is 3.12 pg/mL-200 pg/mL and the sensitivity is 0.78 pg/mL.