Intra-assay Precision (Precision within an assay): CV%<8% | |||||||||||||
Three samples of known concentration were tested twenty times on one plate to assess. | |||||||||||||
Inter-assay Precision (Precision between assays): CV%<10% | |||||||||||||
Three samples of known concentration were tested in twenty assays to assess. | |||||||||||||
To assess the linearity of the assay, samples were spiked with high concentrations of rat LT-B4 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | |||||||||||||
Sample | Serum(n=4) | ||||||||||||
1:1 | Average % | 94 | |||||||||||
Range % | 89-99 | ||||||||||||
1:2 | Average % | 97 | |||||||||||
Range % | 91-103 | ||||||||||||
1:4 | Average % | 97 | |||||||||||
Range % | 92-104 | ||||||||||||
1:8 | Average % | 93 | |||||||||||
Range % | 88-97 |
The recovery of rat LT-B4 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | |||||||||||||
Sample Type | Average % Recovery | Range | |||||||||||
Serum (n=5) | 92 | 87-96 | |||||||||||
EDTA plasma (n=4) | 97 | 93-106 | |||||||||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]()
|
This Rat LT-B4 ELISA Kit was designed for the quantitative measurement of Rat LT-B4 protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 15.6 pg/mL-1000 pg/mL and the sensitivity is 3.9 pg/mL.