Code | CSB-E12948h |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * The sample kit cost can be deducted from your subsequent orders of 96T full size kits of the same analyte at 1/5 per kit, until depleted in 6 months. Apply now |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human Oxyntomodulin in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 89 | ||||
Range % | 83-94 | |||||
1:2 | Average % | 84 | ||||
Range % | 80-89 | |||||
1:4 | Average % | 95 | ||||
Range % | 89-99 | |||||
1:8 | Average % | 92 | ||||
Range % | 88-96 |
The recovery of human Oxyntomodulin spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 95 | 91-99 | ||||
EDTA plasma (n=4) | 98 | 92-103 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
The Human oxyntomodulin ELISA Kit is specifically designed for the detection and quantification of oxyntomodulin in human samples. Oxyntomodulin is a hormone produced in the gastrointestinal tract that has been linked to various metabolic functions, making our ELISA kit an essential tool for researchers in the field of Metabolism.
This ELISA kit is optimized for the detection of oxyntomodulin in human serum, plasma, cell culture supernates, urine, and tissue homogenates. With a detection range of 0.312 ng/mL-20 ng/mL and a sensitivity of 0.078 ng/mL, our kit provides highly accurate and reliable results, allowing you to obtain precise measurements of oxyntomodulin in your samples.
The assay time ranges from 1-5 hours, and only requires a sample volume of 50-100ul, making our ELISA kit a quick and efficient solution for your research needs. The detection wavelength of 450 nm provides clear and reliable readings, while over 5 citations demonstrate the scientific validity and reliability of this kit.