Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human IL-7 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | ||||
1:1 | Average % | 87 | ||||
Range % | 82-94 | |||||
1:2 | Average % | 96 | ||||
Range % | 89-103 | |||||
1:4 | Average % | 92 | ||||
Range % | 85-98 | |||||
1:8 | Average % | 90 | ||||
Range % | 86-95 | |||||
The recovery of human IL-7 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 92 | 85-99 | ||||
EDTA plasma (n=4) | 96 | 89-103 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||
| |||||||
pg/ml | OD1 | OD2 | Average | Corrected | |||
100 | 2.476 | 2.630 | 2.553 | 2.404 | |||
50 | 1.635 | 1.667 | 1.651 | 1.502 | |||
25 | 0.996 | 1.028 | 1.012 | 0.863 | |||
12.5 | 0.590 | 0.623 | 0.606 | 0.458 | |||
6.25 | 0.380 | 0.391 | 0.385 | 0.237 | |||
3.12 | 0.277 | 0.263 | 0.270 | 0.122 | |||
1.56 | 0.222 | 0.214 | 0.218 | 0.069 | |||
0 | 0.152 | 0.145 | 0.149 |
| |||
This ELISA kit provides quantitative measurement of interleukin-7 (IL-7) in human samples, supporting research into lymphocyte development, T-cell homeostasis, and immune reconstitution. Because circulating IL-7 levels are typically low and fluctuate inversely with lymphocyte counts, precise detection across both steady-state and lymphopenic conditions demands a carefully calibrated assay.
Detection Range: The 1.56–100 pg/mL range spans physiological serum IL-7 concentrations in healthy adults (~5–15 pg/mL) through the elevated levels observed during lymphopenia, HIV-associated CD4 depletion, or post-chemotherapy immune recovery. Most serum and plasma samples can be assayed neat, minimizing dilution-related variability.
Sensitivity: At 0.39 pg/mL, the kit detects IL-7 in low-abundance contexts such as unstimulated stromal cell culture supernatants or samples from individuals with intact lymphocyte pools where IL-7 is consumed rapidly and circulates at single-digit pg/mL concentrations.
Sample Compatibility: Validation across serum, plasma, tissue homogenates, and cell culture supernatants enables researchers to correlate systemic IL-7 availability with local production in thymic or bone marrow tissue and in vitro stromal co-culture systems within a single study.
Assay Efficiency: Requiring only 50–100 µL per well and 1–5 hours to complete, the kit accommodates limited clinical specimens from lymphopenic patients and supports same-day data acquisition from time-course experiments.
This kit is well suited for studies of immune reconstitution following hematopoietic stem cell transplantation, HIV pathogenesis, and T-cell acute lymphoblastic leukemia.