| Code | CSB-E04677h-IS |
| Size | 96T,5×96T,10×96T |
| Price | Request a Quote |
| Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
| Have Questions? | Leave a Message or Start an on-line Chat |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human MMP3/STR1 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | ||||
1:40 | Average % | 106 | ||||
Range % | 99-113 | |||||
1:80 | Average % | 100 | ||||
Range % | 93-107 | |||||
1:160 | Average % | 91 | ||||
Range % | 84-97 | |||||
1:320 | Average % | 92 | ||||
Range % | 86-98 | |||||
The recovery of human MMP3/STR1 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 100 | 93-107 | ||||
EDTA plasma (n=4) | 93 | 88-98 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||
| |||||||
pg/ml | OD1 | OD2 | Average | Corrected | |||
2000 | 3.426 | 3.234 | 3.330 | 3.173 | |||
1000 | 1.777 | 1.787 | 1.782 | 1.624 | |||
500 | 0.733 | 0.780 | 0.756 | 0.599 | |||
250 | 0.432 | 0.429 | 0.431 | 0.273 | |||
125 | 0.305 | 0.297 | 0.301 | 0.144 | |||
62.5 | 0.246 | 0.253 | 0.249 | 0.092 | |||
31.25 | 0.174 | 0.184 | 0.179 | 0.022 | |||
0 | 0.160 | 0.155 | 0.157 |
| |||
This ELISA kit provides quantitative measurement of human MMP3 (stromelysin-1) in research contexts where matrix metalloproteinase activity drives tumor invasion, stromal remodeling, and metastatic progression. Because MMP3 concentrations can vary dramatically between baseline circulation and active tumor microenvironment secretion, this kit is designed to capture that full dynamic range across both systemic and tissue-level sampling.
Detection Range: The 31.25–2000 pg/mL range spans reported circulating MMP3 levels in healthy individuals through the elevated concentrations observed in patients with invasive carcinomas. Most serum samples can be assayed neat or with minimal dilution, while tissue homogenates from MMP3-rich tumor stroma may require predilution to fall within range.
Sensitivity: At 15.6 pg/mL, the kit detects MMP3 in low-abundance settings such as early-stage tumor serum samples or conditioned media from epithelial cells undergoing epithelial-mesenchymal transition, where MMP3 upregulation may be subtle.
Sample Compatibility: Validation across serum, plasma, and tissue homogenates (50–100 µL) enables paired designs comparing circulating MMP3 with local expression in resected tumor and adjacent normal tissue, critical for correlating systemic biomarker levels with stromal protease activity.
Assay Efficiency: The 1–5 hour protocol and low sample volume accommodate high-throughput screening of clinical cohorts and conserve limited biopsy-derived homogenates.
This kit supports studies in breast cancer stromal invasion, colorectal tumor progression, and MMP3-driven metastatic niche remodeling.