Code | CSB-E07967r |
Size | 96T,5×96T,10×96T |
Price | Request a Quote or Start an on-line Chat |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * The sample kit cost can be deducted from your subsequent orders of 96T full size kits of the same analyte at 1/5 per kit, until depleted in 6 months. Apply now |
Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<15% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of rat PG-E2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 96 | ||||
Range % | 91-99 | |||||
1:2 | Average % | 92 | ||||
Range % | 87-96 | |||||
1:4 | Average % | 93 | ||||
Range % | 89-98 | |||||
1:8 | Average % | 105 | ||||
Range % | 99-109 |
The recovery of rat PG-E2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 96 | 92-99 | ||||
EDTA plasma (n=4) | 101 | 96-104 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||
|
This Rat PG-E2 ELISA Kit was designed for the quantitative measurement of Rat PG-E2 protein in serum, plasma, tissue homogenates, cell lysates. It is a Competitive ELISA kit, its detection range is 0.5 pg/mL-100 pg/mL and the sensitivity is 0.25 pg/mL.
Sample type: Tissue homogenate(brain)
Sample species: Mouse
Review: One model group differed from the control group, the other two did not. The operation steps of the product are still very simple, and the standards are also prepared with different concentrations.
By Anonymous