Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human MIG/CXCL9 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | ||||
1:1 | Average % | 95 | ||||
Range % | 91-99 | |||||
1:2 | Average % | 87 | ||||
Range % | 83-90 | |||||
1:4 | Average % | 104 | ||||
Range % | 100-108 | |||||
1:8 | Average % | 88 | ||||
Range % | 85-92 |
The recovery of human MIG/CXCL9 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 102 | 97-106 | ||||
EDTA plasma (n=4) | 110 | 107-113 |
| ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
2500 | 2.151 | 2.204 | 2.178 | 2.070 | ||
1250 | 1.647 | 1.602 | 1.625 | 1.517 | ||
625 | 1.045 | 1.104 | 1.075 | 0.967 | ||
312 | 0.701 | 0.731 | 0.716 | 0.608 | ||
156 | 0.411 | 0.425 | 0.418 | 0.310 | ||
78 | 0.253 | 0.256 | 0.255 | 0.147 | ||
39 | 0.150 | 0.157 | 0.154 | 0.046 | ||
0 | 0.107 | 0.109 | 0.108 |
|