| Code | CSB-E08412r |
| Size | 96T,5×96T,10×96T |
| Price | Request a Quote |
| Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
| Have Questions? | Leave a Message or Start an on-line Chat |
| Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<10% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of rat sIgA in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:100 | Average % | 102 | ||||
| Range % | 97-106 | |||||
| 1:200 | Average % | 90 | ||||
| Range % | 86-96 | |||||
| 1:400 | Average % | 97 | ||||
| Range % | 92-103 | |||||
| 1:800 | Average % | 88 | ||||
| Range % | 84-92 | |||||
| The recovery of rat sIgA spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 84 | 80-88 | ||||
| EDTA plasma (n=4) | 95 | 91-99 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This Rat sIgA ELISA Kit was designed for the quantitative measurement of Rat sIgA protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 0.78 ng/mL-50 ng/mL and the sensitivity is 0.195 ng/mL .
Sample type: Serum
Sample species: Rat
Sample dilution: No dilution
Review: I used CSB-E08412R to detect the content of HMGB1 in colon and lung tissue homogenates. The product is convenient and fast, with high sensitivity. The experimental results are reliable, the detection results are reliable, and the trend is obvious. It is recommended to use.