Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-6 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | ||||
1:1 | Average % | 93 | ||||
Range % | 88-99 | |||||
1:2 | Average % | 90 | ||||
Range % | 86-96 | |||||
1:4 | Average % | 95 | ||||
Range % | 88-101 | |||||
1:8 | Average % | 92 | ||||
Range % | 86-98 | |||||
The recovery of mouse IL-6 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 90 | 85-96 | ||||
EDTA plasma (n=4) | 99 | 93-105 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||
| |||||||
pg/ml | OD1 | OD2 | Average | Corrected | |||
250 | 2.585 | 2.634 | 2.610 | 2.439 | |||
125 | 1.736 | 1.667 | 1.701 | 1.531 | |||
62.5 | 0.956 | 1.018 | 0.987 | 0.816 | |||
31.25 | 0.592 | 0.621 | 0.606 | 0.436 | |||
15.6 | 0.408 | 0.411 | 0.409 | 0.239 | |||
7.8 | 0.284 | 0.273 | 0.279 | 0.108 | |||
3.9 | 0.224 | 0.214 | 0.219 | 0.048 | |||
0 | 0.166 | 0.175 | 0.170 |
| |||
This kit provides quantitative measurement of interleukin-6 (IL-6) in mouse samples, supporting immunology research focused on this pleiotropic cytokine's roles in inflammation, acute-phase responses, and immune regulation. Because IL-6 levels can shift dramatically — from low picogram concentrations at baseline to elevated levels during infection or chronic inflammation — the kit is designed to capture this dynamic range across diverse experimental models.
Detection Range: The 3.9–250 pg/ml range spans basal circulating IL-6 levels in healthy mice (often <20 pg/ml) through the elevated concentrations seen in LPS challenge, collagen-induced arthritis, or tumor-bearing models, minimizing the need for serial dilutions in most acute inflammation studies while still resolving subtle baseline differences.
Sensitivity: At 0.975 pg/ml, the kit detects IL-6 in unstimulated cell culture supernates and serum from naïve mice, where concentrations frequently fall below 5 pg/ml — critical for establishing true baseline values or measuring early innate immune activation.
Sample Compatibility: Validation across serum, plasma, tissue homogenates, and cell culture supernates enables researchers to correlate systemic IL-6 levels with local tissue production and in vitro macrophage or T cell stimulation data within a single assay platform.
Assay Efficiency: Requiring only 50–100 µl per well and 1–5 hours to complete, the kit accommodates limited-volume samples from retro-orbital bleeds or primary splenocyte cultures while supporting same-day data acquisition.
This kit is particularly suited for studies in sepsis modeling, colitis-associated tumorigenesis, and Th17 differentiation pathways.