Code | CSB-E08664b |
Size | 96T,5×96T,10×96T |
Price | Request a Quote |
Trial Size |
24T ELISA Kit Trial Size (Only USD$150/ kit) * Sample kit cost can be deducted as a $30 credit for each 96-assay kit of the same analyte and brand you subsequently purchase within six months until depleted. More details >> Interested in a trial size? Please leave a message below.
|
Have Questions? | Leave a Message or Start an on-line Chat |
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||||||||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||||||||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||||||||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||||||||||||
| Intra-Assay Precision | Inter-Assay Precision | ||||||||||||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 | ||||||||||
n | 20 | 20 | 20 | 20 | 20 | 20 | ||||||||||
Mean(μg/ml) | 5.265 | 5.989 | 5.270 | 5.475 | 5.510 | 4.853 | ||||||||||
SD | 0.038 | 0.033 | 0.038 | 0.041 | 0.044 | 0.048 | ||||||||||
CV(%) | 6.463 | 5.546 | 6.179 | 6.732 | 7.237 | 7.643 |
To assess the linearity of the assay, samples were spiked with high concentrations of Bovine Alb in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | |||||||||||||||||
Sample | Serum(n=4) | ||||||||||||||||
1:2000 | Average % | 96 | |||||||||||||||
Range % | 89-113 | ||||||||||||||||
1:4000 | Average % | 92 | |||||||||||||||
Range % | 86-107 | ||||||||||||||||
1:8000 | Average % | 98 | |||||||||||||||
Range % | 85-108 | ||||||||||||||||
1:16000 | Average % | 90 | |||||||||||||||
Range % | 83-98 | ||||||||||||||||
The recovery of Bovine Alb spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | |||||||||||||||||
Sample Type | Average % Recovery | Range% | |||||||||||||||
Serum (n=5) | 93 | 88-95 | |||||||||||||||
EDTA plasma (n=4) | 93 | 89-97 | |||||||||||||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||
μg/ml | OD1 | OD2 | Average | ||||||||||||||||
80 | 0.309 | 0.289 | 0.299 | ||||||||||||||||
20 | 0.443 | 0.423 | 0.433 | ||||||||||||||||
5 | 0.579 | 0.568 | 0.573 | ||||||||||||||||
1.25 | 0.857 | 0.848 | 0.853 | ||||||||||||||||
0.313 | 1.504 | 1.498 | 1.501 | ||||||||||||||||
0 | 2.242 | 2.238 | 2.240 | ||||||||||||||||
The Bovine Albumin (Alb) ELISA Kit is designed for the quantitative measurement of bovine albumin in various sample types, including serum, plasma, and tissue homogenates. This kit utilizes a competitive detection mechanism to accurately quantify bovine albumin. It features a detection range of 0.313 μg/mL to 80 μg/mL, with a sensitivity of 0.148 μg/mL, making it highly sensitive and reliable for research applications. The assay time ranges from 1 to 5 hours, and the required sample volume is between 50 to 100 microliters. Detection is performed at a wavelength of 450 nm, ensuring precise and reproducible results. This ELISA kit is an essential tool for researchers studying bovine albumin in biological samples.